DO NOT USE:
•
solutions containing phosphoric acid (Foamy
aromatic solvents (such as naphtha, paint thinner),
chlorinated solvents
Toluene), ammonia, acetone, benzene, xylene or alcohol,
other than as specified below.
•
hard or pointed objects to clean any part of instrument.
All recommended solutions must be diluted per the
manufacturer's recommendation. Acceptable cleaning
solutions are:
2% Glutaraldehyde in water
2% Phenols in water (O-Syl 1:128, Pheno-Cen 1:256, Vesphene)
10% bleach solution (1 part bleach to 9 parts water)
70% Isopropyl Alcohol
CaviCide
Compublend II
Envirocide
Hydrogen Peroxide 3%
Mild detergent (such as Manu-Klenz)
Quaternaries 1:256 and 1:512
Sani-Cloth HB
Sani-Cloth Plus
Super Sani-Cloth
Warm water
WEX-CIDE
1.
Keep instrument upright and do not allow any part of
instrument to become saturated with or submersed in fluid
during cleaning operation.
2.
Use a soft cloth dampened with warm water and a mild
nonabrasive cleaning solution to clean all exposed
surfaces. A soft-bristled brush may be used to clean hard
to reach and narrow areas. For sanitizing or antibacterial
treatment, use
10%
3.
Use a soft cloth dampened with water to rinse off cleaning
solution.
Alaris
®
System (with v9 Model 8000) DFU
Appendix Section
Cleaning
(such as Trichloroethane,
bleach solution and water.
WARNING
) ,
Q&A
To prevent an electrical hazard:
• Turn the instrument off and
,
MEK
unplug the power cord from
power before cleaning.
• Do not spray fluids directly onto
the instrument or into the
connectors.
• Do not steam autoclave,
sterilize, immerse the instrument
in fluids, or allow fluids to enter
the instrument case.
CAUTION
The solutions/solvents identified
as
NOT
surfaces of the instrument.
Maintenance
IUI
E
to be used can damage the
Maintenance
AC
t
O
A-1