Disabling Pca Pause Alarm - Cardinal Health Alaris Directions For Use Manual

Supports guardrails suite mx with guardrails point-of-care software and v9 operating system software. with pc unit, model 8000.
Table of Contents

Advertisement

1.
From Main Display press
OPTIONS
2.
Press
Press PCA Pause Limits soft key.
3.
DISABLE SPO2
4.
Press
appropriate.
CONFIRM
5.
Press
START
6.
Press
soft key.
7.
To enable PCA Pause feature, follow steps
ENABLE SPO2
and press
appropriate.
NOTES:
Disabling
p
S
O
2
PCA Pause feature only, without interrupting monitoring
functionality.
Once disabled, the alarm limits are grayed out and are not
editable.
3-36
Programming

Disabling PCA Pause Alarm

CHANNEL SELECT
key.
DISABLE ETCO2
or
soft key.
ENABLE ETCO2
or
or
t
from this screen discontinues the
E
CO
2
.
soft key, as
-
above
1
3
soft key, as
Alaris
A
PCA Pause Limits
PCA
SPO2/ETCO2
PAUSE LIMITS
ALARM LIMITS
97
88
%SPO2
LOW
5
_4
RR
LOW
>Enter a Pause Limit
Between 2 - 4
DISABLE
CONFIRM
SPO2
®
System (with v9 Model 8000) DFU
PCA Module Section
DISABLE
ETCO2

Advertisement

Table of Contents
loading

Table of Contents