References throughout this procedure to specific drugs and
drug doses are for illustration purposes only. Reference
specific drug product labeling for information concerning
appropriate administration techniques and dosages.
See "General Information", "Features and Displays" and the
PC Unit Section of this DFU for information about:
•
Displays
•
Operating Features, Controls, Indicators
Selecting Syringe Type and Size
At the start of an infusion program, the system prompts user to
select and confirm the syringe type and size.
CHANNEL SELECT
1.
Press
position.
®
Alaris
System (with v9 Model 8000) DFU
PCA Module Section
Preparing Infusion
PROGRAM
key. Key must be in
Programming
WARNING
Ensure the displayed syringe
manufacturer and size correctly
identifies the installed syringe.
Mismatches may cause an under-
infusion or over-infusion to the
patient that could result in serious
injury and/or death. For a list of
compatible syringes, see "General
Information", "Compatible Syringes".
If the installed syringe is displayed
and selected, but is not recognized,
servicing is required (see "Service
Information" in "Appendix" Section of
this DFU).
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
PCA Module
Programming
3-7