Infusion - NO
The following procedures should be used only to set up a
Basic Infusion. To program an infusion using Guardrails
Drugs, see "Primary Infusion - With Guardrails
Protection".
The illustrations in this procedure assume:
•
Mode (Syringe Module), Drug Calculation, Dynamic
ALL
Pressure Display, Profiles, and Volume Duration
configurable settings are enabled.
•
Syringe Module) and Delay Options configurable
NEOI (
settings are disabled.
1.
Press
CHANNEL SELECT
2.
Press Basic Infusion soft key.
•
Infusion Setup screen appears.
3.
Start applicable infusion, as described in following
procedures.
Rate/Volume Infusion
Volume/Duration Infusion
NOTE:
If Delay Options is enabled, the
.
DELAY OPTIONS
Rate / Volume and Volume / Duration
Infusions
See "Primary Infusion - With Guardrails
"IV Fluid Infusion" procedure.
®
Alaris
System (with v9 Model 8000) DFU
Pump and Syringe Modules Section
Guardrails
®
Suite MX Protection (Continued)
Basic Infusion
key.
soft key becomes
PAUSE
®
Suite MX Protection",
When the pressure sensing disc
®
Suite MX
is not being used and an occlusion
occurs, there is a risk of infusing
pressurized buildup of infusates
upon correction of the occlusion.
To avoid an inadvertent bolus,
relieve the pressure before
restarting the infusion.
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
WARNING
Programming
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
2-51