Tubing Set; Visual Inspection; Cleaning; Leak Test - Cardinal Health Kendall SCD 700 Manual

Table of Contents

Advertisement

Section 2 – Instructions for Customer Returned Units (continued)

TUBING SET

VISUAL INSPECTION

Inspect the tubing sets for missing, broken or cracked components. If any components are
missing, broken or cracked, discard the tubing set.
Inspect the tubing set for condensation in the tubing. If condensation is present in the tubing,
discard the tubing set.
Inspect the printing on the tubing set connectors. If any connectors contain incomplete
characters, discard the tubing set.

CLEANING

Clean the customer returned tube set as per the Kendall SCD™ 700 system Operation and
Service Manual.
All visible extraneous markings or materials should be removed by the cleaning step. If
extraneous markings or materials are not removed, discard the tubing set.

LEAK TEST

Inspect the Leak Test Fixture to ensure that it is in working order and that the pressure gauge
has a valid calibration date.
Ensure that the Leak Test Fixture is connected to a regulated air source. The input air should
be at pressure between 3 psi and 4 psi. See sketch below.
Connect the Tubing Set to the connectors on the test fixture. See pictures below.
CONNECT TO AIR
7