Secondary Infusion
To continue programming, press Yes soft key.
4.
OR
To change selection, press No soft key.
If Yes was selected and facility has defined a Clinical
•
Advisory for that drug, a message appears. To
indicate information has been noted and continue
programming, press
If Yes was selected to continue programming, drug
•
amount and diluent volume (if defined in Drug Library)
are automatically entered for selected drug.
•
If selected drug had "_ _ / _ _ mL" concentration, drug
amount and diluent volume need to be entered.
If selected drug is not weight-based, Not Used
•
PATIENT WEIGHT
displays in
•
If hospital/facility practice guidelines identify selected
drug as weight-based, prompt for a patient weight in
kilograms or
which reflects use of Methotrexate).
5.
Verify parameters are correct and press
confirm.
•
If programmed total dose drug amount is outside Soft
Limit for that care area, an audio alert sounds and
a visual prompt appears before programming can
continue. If Yes soft key is pressed, programming
continues; if No soft key is pressed, infusion needs to
be reprogrammed.
•
If programmed total dose drug amount is outside
Hard Limit for that care area, an audio alert sounds
and a visual prompt appears before programming can
continue. Infusion needs to be reprogrammed.
-- Continued on Next Page --
®
Alaris
System (with v9 Model 8000) DFU
Pump and Syringe Modules Section
- With Guardrails
(
Pump Module
Infusion (Continued)
CONFIRM
soft key.
field.
appears (as in illustrated example,
BSA
Suite MX Protection
®
) (Continued)
NEXT
soft key to
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
A
Guardrails Drug Setup
Peds Oncology
Methotrexate
Yes
_ _ _ mg/_ _ _mL was
No
selected.
Is this correct?
2
BSA
DOSING
mg/m
UNITS
>Press Yes or No
A
Guardrails Drug Setup
Methotrexate
DRUG
6840
mg
AMOUNT
20
DILUENT
mL
VOLUME
_ _ _
2
BSA
m
_ _ _
mg/m
2
DOSE =
[Conc]: 1.7 mg/mL
>Enter BSA
DRUG
PRIMARY
LIBRARY
Programming
2-47