Preparing Syringe and Administration Set (
Priming - Using Options Menu (continued)
Administration Set With Pressure Sensing
Disc (Continued)
Press Prime Set with Syringe soft key.
4.
•
If pressure sensing disc was not removed prior to
pressing Prime Set with Syringe soft key, a pressure
sensing disc removal prompt displays.
5.
Invert pressure sensing disc so that patient side is up.
6.
Hold pressure sensing disc between
®
Alaris
System (with v9 Model 8000) DFU
Pump and Syringe Modules Section
Syringe Module
fingers.
2
FU
RATE (mL/h)
)
(Continued)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
A
Channel Options 1 of 2
Guardrails Drugs
Guardrails IV Fluids
Multidose
Pressure Limit - 550 mmHg
Prime Set with Syringe
>Select an Option or EXIT
EXIT
DOWN
A A A
It is recommended that the
Pressure Sensing Disc be
removed during priming.
Press CONFIRM and
continue priming.
>Press CONFIRM
CONFIRM
Patient Side
Getting Started