Secondary Infusion
The following procedure should be used only when:
• drug to be infused is listed in Drug Library,
• primary infusion is running, and
• a check valve administration set is being used.
To program a primary infusion, see "IV Fluid Infusion"
procedure. To program a basic infusion, see "Infusion -
®
Guardrails
Suite MX Protection" procedure.
CHANNEL SELECT
1.
Press
SECONDARY
2.
Press
3.
Press soft key next to desired drug.
If applicable, an optional hospital-defined therapy
or clinical indication for delivery of this infusion may
appear. Different limits can be defined for same drug
with different therapeutic indications.
If applicable, a weight-based, non weight-based, or
-based option for delivery of this infusion may
BSA
appear.
If applicable, multiple concentration listings for delivery
of this infusion may appear.
2-46
Programming
- With Guardrails
(
) (Continued)
Pump Module
Infusion
key.
soft key.
Suite MX Protection
®
NO
®
Alaris
System (with v9 Model 8000) DFU
Pump and Syringe Modules Section
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
Guardrails Fluid Setup
A
0.9% NaCI
PRIMARY INFUSION
40
mL/h
RATE
1000
VTBI
mL
>Press START
DELAY
VOLUME
SECOND-
START
OPTIONS
DURATION
ARY
A
A
Guardrails Drugs
Peds Oncology
Amifostine
A-E
Azathioprine
F-J
Bleomycin
K-O
Cytarabine
P-T
Daunorbicin
U-Z
>Select Drug
DRUG
PAGE
EXIT
DOWN
CALC