General Information; Warnings And Cautions - Cardinal Health Alaris Directions For Use Manual

Supports guardrails suite mx with guardrails point-of-care software and v9 operating system software. with pc unit, model 8000.
Table of Contents

Advertisement

WARNINGS
• The nurse call feature should not be used as the primary
source of alarm notification. The audible and visual
alarms of the Alaris
clinical signs and symptoms, are the primary sources
providing notification that an alarm condition exists.
• Before each use, ensure the nurse call cable is fully
inserted into the jack, and test the operation of the nurse
call system by causing an alarm and verifying proper nurse
call operation.
CAUTIONS
• To assure compliance with Electromagnetic
Compatibility Standards (
option must only be used with a properly configured
hospital interface cable assembly. The approved hospital
nurse call interface cable consists of an
cable with an attached Fair-Rite Suppressor Core
24-660
Assembly. The Fair-Rite part is included with the Nurse
Call Accessory. Reference the cable assembly instructions,
included with the Nurse Call Accessory, for additional
information.
• When moving the PC Unit, ensure the nurse call cable is
disconnected from the nurse call jack on the PC Unit and
the fluid seal covers the jack opening.
• Carefully route cabling to reduce the possibility of
entanglement that could result in a tripping hazard.
®
Alaris
System (with v9 Model 8000) DFU
Nurse Call Accessory Section

Warnings and Cautions

®
System, used in conjunction with
), the nurse call
IEC 60601-1-2
MCM
Electronics

General Information

General Information
7-3