Syringe Adjustment Alarms (Continued)
When problem is corrected, press
Alarm
Check Syringe
Drive Not Engaged
2-120
Troubleshooting and Maintenance
Alarms, Errors, Messages (Continued)
Alarms (Continued)
CONFIRM
Plunger grippers opened during
infusion and then closed. Infusion
stops on affected module.
Syringe barrel clamp opened
during infusion and then closed.
Infusion stops on affected module.
Syringe plunger not captured
while in idle state. System alarms
after
30
potential siphoning condition.
Drive system disengaged during
operation.
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
soft key.
Meaning
seconds, to indicate
A A A
A
Drug Calculation
Heparin
CONTINUOUS INFUSION
Syringe Installation
plunger
flange
>Press CONFIRM
Response
Securely lock plunger grippers,
CHANNEL SELECT
press
reselect syringe.
Securely lock syringe barrel clamp
RESTART
and press
Check for potential siphoning.
Ensure administration set clamp
(roller/slide) is in closed position.
Securely lock plunger grippers over
syringe plunger.
Open and close plunger grippers
and syringe barrel clamp. Ensure
syringe is properly installed.
Alaris
®
System (with v9 Model 8000) DFU
Pump and Syringe Modules Section
lever
clamp
CONFIRM
key, and
key.