Cardinal Health Alaris Directions For Use Manual page 80

Supports guardrails suite mx with guardrails point-of-care software and v9 operating system software. with pc unit, model 8000.
Table of Contents

Advertisement

Preparing Syringe and Administration Set (
Priming - Using Options Menu (continued)
Administration Set With No Pressure
Sensing Disc
OPTIONS
1.
Press
key.
Press Prime Set with Syringe soft key.
2.
PRIME
3.
Press and hold
is complete.
PRIME
4.
Release
soft key.
5.
To return to main screen, press
NOTES:
When manually priming (per hospital/facility protocol) and an
administration set having a pressure sensing disc is in use,
depress the disc between
uphill (distal end of pressure sensing disc/tubing pointing
upward).
Fluid is delivered during priming only while the
key is pressed. Each press of the
up to
mL of priming fluid per continuous press. To deliver
2
additional amounts, press the
Volume used during priming is displayed but not added to
or
.
VTBI
VI
2-16
Getting Started
soft key until fluid flows and priming
EXIT
soft key.
fingers while priming and prime
2
PRIME
PRIME
soft key delivers
PRIME
soft key again.
)
Syringe Module
(Continued)
soft
®
Alaris
System (with v9 Model 8000) DFU
Pump and Syringe Modules Section
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
A
Channel Options 1 of 2
Guardrails Drugs
Guardrails IV Fluids
Multidose
Pressure Limit - High
Prime Set with Syringe
>Select an Option or EXIT
PAGE
EXIT