Infusion - NO
Continuous Infusion - Drug Calculation (Continued)
NOTES:
Do not enter a patient weight if weight is not used in the
calculation.
Pump Module: At rates less than
displayed to two decimal places, and the
and is displayed to two decimal places.
Pump Module: In the Drug Calculation mode, the system
infuses at the calculated rate rounded to the nearest one-
hundredth of a mL per hour (as displayed on the programming
screen). The rate shown in the Rate Display is rounded to the
nearest one-tenth of a mL per hour.
Syringe Module: Starting the infusion before attaching the
administration set to the patient minimizes any potential bolus
that can be released from pressure built up in the set due to
normal syringe loading and priming.
1.
Set up infusion as described in "Continuous Infusion -
Drug Calculation" procedure, but do not start infusion.
BOLUS
2.
Press
soft key.
3.
To enter bolus dose, use numeric data entry keys.
After a bolus dose and weight (if used) are entered,
bolus
and concentration [conc] alternate in Main
VTBI
Display.
4.
Press soft key next to appropriate unit of measure for
dose.
5.
To enter bolus duration, use numeric data entry keys.
TOTAL DOSE
6.
Verify parameters are correct and press
2-54
Programming
Guardrails
Suite MX Protection (Continued)
®
mL/h, the rate is
10
Bolus Dose
INFUSE AT
alternates with
can be entered
VTBI
rate.
START
soft key.
A A A
Drug Calculation
BOLUS DOSE
_ 2000
DOSE
PATIENT
WEIGHT
DURATION
[Conc]: 5000 mcg/mL
>Select the Desired Dosing
Units
SETUP
®
Alaris
System (with v9 Model 8000) DFU
Pump and Syringe Modules Section
FU
FU
RATE (mL/h)
RATE (mL/h)
CHANNEL
SELECT
CHANNEL
SELECT
PAUSE
PAUSE
CHANNEL
OFF
CHANNEL
OFF
RESTART
RESTART
mcg
mcg/kg
mg
mg/kg
CONT-
INUOUS