Programming an Infusion (Continued)
6.
Perform following steps:
a.
Load syringe and administration set (see "Getting
Started", "Loading Syringe and Administration Set").
b.
Select and confirm syringe type and size (see
"Selecting Syringe Type and Size").
7.
Press soft key next to desired drug.
•
Drug/Concentration screen appears.
8.
Press soft key next to desired concentration.
•
Drug/Concentration confirmation screen appears.
•
To view additional drugs/concentrations, press
UP
PAGE DOWN
and
•
Facility may choose to prepopulate standard drug
concentrations, or leave an open entry (_ _ / _ _ mL)
and allow clinician to enter drug amount and diluent
volume.
Confirm drug and concentration selection and press Yes
9.
soft key. To change selection, press No soft key.
If Yes was selected and facility has defined a Clinical
•
Advisory for that drug, a message appears. To
continue programming, press
10. Verify parameters are correct and press
confirm.
11. Prime syringe using Prime feature, if desired.
®
Alaris
System (with v9 Model 8000) DFU
PCA Module Section
Preparing Infusion (Continued)
soft keys.
CONFIRM
NEXT
PAGE
soft key.
soft key to
A
Guardrails Drugs
Med Surg
Morphine
A-E
Meperidine
F-J
K-O
P-T
U-Z
>Select Drug
EXIT
Guardrails Drugs
A
Med Surg
Morphine
1 mg/mL
5 mg/mL
10 mg/mL
>Select Concentration
DRUG
EXIT
LIBRARY
Programming
3-11