Primary Infusion
Intermittent Infusion (Continued)
5.
entry:
VTBI
Pump Module:
volume of infusion. To change
key and use numeric data entry keys.
Syringe Module:
VTBI ALL
When
inactive and estimated available volume in syringe
displays. When
value can be entered.
VTBI ALL
When
pressed, estimated available volume in syringe
displays.
6.
If an optional hospital-defined and editable starting value
for intermittent duration is not already entered, enter
duration or rate, as follows:
To enter duration, press
numeric data entry keys (rate value is calculated and
displayed).
To enter rate, press
numeric data entry keys.
7.
Verify parameters are correct and press
If programmed duration is outside Soft Limit for that
care area, an audio alert sounds and a visual prompt
appears before programming can continue. If Yes soft
key is pressed, programming continues; if No soft key
is pressed, infusion needs to be reprogrammed.
If programmed duration is outside Hard Limit for that
care area, an audio alert sounds and a visual prompt
appears before programming can continue. Infusion
needs to be reprogrammed.
If a Soft Limit is overridden, G icon displays. When G
soft key is pressed, all applicable out-of-range limits
are listed.
2-38
Programming
- With Guardrails
is prepopulated with diluent
VTBI
, press
VTBI
VTBI
is enabled,
soft key is
VTBI
soft key is pressed, a new
VTBI
is disabled and
DURATION
soft key and use
RATE VOLUME
soft key and use
START
Suite MX Protection (Continued)
®
VTBI
soft
soft key is
soft key.
Alaris
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
Guardrails Drug Setup
A
Methotrexate
PRIMARY INFUSION
_ _ _
RATE =
mL/h
ALL
VTBI
_ _ _ _
DURATION
h
1200
DOSE =
mg/m
[Conc]: 136.8 mg/mL
>Select DURATION
RATE
SETUP
VOLUME
Guardrails Drug Setup
A
Methotrexate
PRIMARY INFUSION
4
RATE =
mL/h
ALL
VTBI
6:00
DURATION
hh:mm
1200
DOSE =
mg/m
[Conc]: 136.8 mg/mL
>Press START
OR
DELAY
RATE
SETUP
OPTIONS
VOLUME
®
System (with v9 Model 8000) DFU
Pump and Syringe Modules Section
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
Available:
(49.2 mL)
min
2
Available:
(49.2 mL)
2
START