Cardinal Health Alaris Directions For Use Manual page 100

Supports guardrails suite mx with guardrails point-of-care software and v9 operating system software. with pc unit, model 8000.
Table of Contents

Advertisement

Primary Infusion
Intermittent Infusion (Continued)
2.
Press soft key next to desired drug.
If applicable, an optional hospital-defined therapy
or clinical indication for delivery of this infusion may
appear. Different limits can be defined for same drug
with different therapeutic indications.
If applicable, a weight-based, non weight-based, or
-based option for delivery of this infusion may
BSA
appear.
If applicable, multiple concentration listings for delivery
of this infusion may appear.
To continue programming, press Yes soft key.
3.
OR
To change selection, press No soft key.
If Yes was selected and facility has defined a Clinical
Advisory for that drug, a message appears. To
indicate information has been noted and continue
programming, press
If Yes was selected to continue programming, drug
amount and diluent volume (if defined in Drug Library)
are automatically entered for selected drug.
If selected drug had "_ _ / _ _ mL" concentration, drug
amount and diluent volume need to be entered.
If selected drug is not weight-based, Not Used
PATIENT WEIGHT
displays in
-- Continued on Next Page --
2-36
Programming
- With Guardrails
Suite MX Protection (Continued)
®
CONFIRM
soft key.
field.
A
Guardrails Drugs
A
Peds Oncology
Amifostine
Azathioprine
Bleomycin
Cytarabine
Daunorbicin
>Select Drug
BASIC
PRIMARY
SEC
A
Guardrails Drug Setup
Peds Oncology
Methotrexate
_ _ _ mg/_ _ _mL was
selected.
Is this correct?
2
DOSING
mg/m
UNITS
>Press Yes or No
®
Alaris
System (with v9 Model 8000) DFU
Pump and Syringe Modules Section
FU
FU
RATE (mL/h)
RATE (mL/h)
CHANNEL
SELECT
CHANNEL
SELECT
PAUSE
PAUSE
CHANNEL
OFF
CHANNEL
OFF
RESTART
RESTART
A-E
F-J
K-O
P-T
U-Z
PAGE
DOWN
Yes
No
BSA

Advertisement

Table of Contents
loading

Table of Contents