Viewing and Clearing Volume Infused

1.
To view volume infused, press
from Main Display.
Total volume infused, and time and date volume infused
was last cleared, is displayed for each channel.
2.
To clear volume infused:
If only selected channels are to be cleared, press soft
key next to applicable channel(s) and press
CHANNEL
soft key.
If all channels are to be cleared, press
key.
3.
To return to main screen, press
NOTES:
Date format is year-month-day.
If no key is pressed, main screen appears after
Clearing volume infused on a PCA Module does not clear
patient history.
The PCA Pause Protocol is an optional, hospital-configurable
feature that is intended to align with the healthcare facility's
current protocol for patient monitoring during PCA Therapy. All
programming, data entry and validation of PCA Pause Protocol
parameters are performed by a healthcare professional
according to hospital-defined protocol/procedure or a
physician's order.
3-32
Programming
Infusion Modes (Continued)
VOLUME INFUSED
MAIN SCREEN

PCA Pause Protocol Feature

soft key
CLEAR
CLEAR ALL
soft
soft key.
seconds.
30
Alaris
Volume Infused
TOTAL VOLUME (mL)
A
401.10
B
42.50
C
478.10
D
789.10
>Select Channels to Clear
or Press CLEAR ALL
CLEAR
PRI/SEC
ALL
VOLUME
®
System (with v9 Model 8000) DFU
PCA Module Section
LAST
CLEARED
08:00
1998-3-10
07:30
1998-3-11
08:00
1998-3-10
12:00
1998-3-10
MAIN
SCREEN