Cardinal Health Alaris Directions For Use Manual page 109

Supports guardrails suite mx with guardrails point-of-care software and v9 operating system software. with pc unit, model 8000.
Table of Contents

Advertisement

Secondary Infusion
1.
Open secondary administration set package, remove set
and close clamp.
2.
Insert administration set spike into prepared fluid container
and hang secondary container, following accepted
hospital/facility procedure.
3.
Fill drip chamber to
4.
Open secondary administration set clamp and prime set.
Close clamp.
5.
Attach secondary administration set to upper injection site
on primary set.
6.
Using hanger provided with secondary administration
set, lower primary fluid container to height indicated in
following illustrations.
Hanger
Primary
Container
®
Alaris
System (with v9 Model 8000) DFU
Pump and Syringe Modules Section
- With Guardrails
(
) (Continued)
Pump Module
Setup
/
full.
2
3
Secondary Container
Mini Bag
at least 9½"
Suite MX Protection
®
WARNINGS
• The clamp on the secondary
administration set must be
opened. If the clamp is not
opened, the fluid is delivered
from the primary container.
• The secondary administration
set must be primed prior to
beginning the secondary infusion.
Primary
Container
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
Secondary Container
Bottle
at least 9½"
2-45
Programming