Preparing Administration Set (
Loading (Continued)
c.
Press safety clamp fitment into recess below
mechanism.
d.
Using a finger tip, firmly push tubing toward back of
Air-in-Line (
AIL
4.
Close door and latch, as follows:
a.
Close door and hold in a closed
position by grasping door and
instrument case with one hand.
b.
Gently lower latch.
Safety clamp device is automatically
disengaged.
5.
Open roller clamp.
6.
Verify no fluid is flowing through drip chamber.
NOTE:
"Safety clamp" is referred to on the device as "Flo-Stop".
2-4
Getting Started
) Detector.
) (Continued)
Pump Module
To reduce the potential for nuisance
AIL
inserted in
• Do not touch the administration
• To prevent a potential free-flow
Alaris
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
CAUTION
alarms, ensure tubing is fully
Detector.
AIL
Safety Clamp Fitment
Push tubing toward
back of AIL Detector.
WARNINGS
set while closing the door. Failure
to follow this instruction may
result in infusion rate inaccuracy.
condition, ensure no extraneous
object (for example, bedding,
tubing, glove) is enclosed or
caught in the Pump Module door.
®
System (with v9 Model 8000) DFU
Pump and Syringe Modules Section