Cardinal Health Alaris Directions For Use Manual page 155

Supports guardrails suite mx with guardrails point-of-care software and v9 operating system software. with pc unit, model 8000.
Table of Contents

Advertisement

Features and Definitions (Continued)
Syringe Module (Continued)
Auto Syringe Size Identification
Back Off
Fast Start
Infusion Complete
Near End of Infusion (NEOI)
Occlusion Pressure
Pressure Sensing Disc
Pressure Tracking
Alaris
®
System (with v9 Model 8000) DFU
Pump and Syringe Modules Section
Features and Displays (Continued)
System automatically detects syringe size and narrows down
syringe selection list.
This feature is only available when administration set in use
has a pressure sensing disc. When enabled, motor reverses
plunger movement during an occlusion until pressure returns to
preocclusion levels, automatically reducing bolus flow.
When Fast Start is enabled and an administration set having a
pressure sensing disc is used, instrument runs at an increased
rate when an infusion is first started, taking-up any slack in drive
mechanism.
An alert is given when current infusion is complete and
reached zero.
Allows an alert to be configured to sound anywhere from
minutes before infusion is complete. Alert occurs at configured
60
time or when
A complete range of downstream occlusion detection options is
provided.
• With pressure sensing disc: Downstream occlusion alarm
threshold is selectable between
increments.
• Without pressure sensing disc: Downstream occlusion alarm
threshold can be set to low, medium, or high.
When installed, pressure sensing disc significantly improves
instrument's pressure sensing capabilities for a faster occlusion
detection time, and makes following features available:
Auto Pressure
Back-Off
Customizable Pressure Alarm Settings (see "Occlusion Pressure")
Fast Start
Pressure Tracking
Dynamic current pressure display is only available when pressure
sensing disc is inserted.
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
of
remains, whichever comes later.
25%
VTBI
and
mmHg, in
25
1000
General Information
has
VTBI
to
1
mmHg
1
2-91

Advertisement

Table of Contents
loading

Table of Contents