Changing Syringe and Restoring Infusion (Continued)
2.
To restart infusion using restored parameters, press
RESTORE
soft key and continue with next step.
OR
To start a new infusion, select drug from Drug Library and
follow steps for "Infusion Modes".
3.
Verify restored drug/concentration. Press
4.
Prime administration set (see "Preparing Infusion",
"Priming").
5.
For restored parameters, verify parameters are valid and
CONFIRM
press
6.
Close and lock security door.
7.
Verify programming parameters on summary screen are
correct and press
NOTES:
If drug and/or drug concentration is different from previous
syringe, attach and prime new administration set.
To change a restored parameter:
a. Press applicable soft key.
b. Enter desired parameter using numeric data entry keys.
CONFIRM
c. Press
3-30
Programming
Infusion Modes (Continued)
soft key.
START
soft key.
soft key.
NEXT
soft key.
Alaris
Guardrails Drugs
A
Med Surg
Morphine
Meperidine
>Select Drug
EXIT
RESTORE
A
Second Nurse Summary
Morphine
2
PCA Dose:
mg
Lockout
10
minutes
Interval:
25
Max Limit:
mg/4 h
[Conc]: 1 mg/mL
>Press START
PRO-
PAUSE
GRAM
®
System (with v9 Model 8000) DFU
PCA Module Section
A-E
F-J
K-O
P-T
U-Z
START