Priming - Cardinal Health Alaris Directions For Use Manual

Supports guardrails suite mx with guardrails point-of-care software and v9 operating system software. with pc unit, model 8000.
Table of Contents

Advertisement

Selecting Syringe Type and Size (Continued)
2.
Press soft key next to installed syringe type and size.
If installed syringe is not listed, press
soft key and select syringe from list.
Selection is highlighted.
soft key is activated.
CONFIRM
3.
To accept, press
Drug Library screen displays.
NOTE:
The system automatically detects the syringe size, and lists
syringe types and sizes that most closely match the installed
syringe. If the syringe is not recognized, Syringe not
recognized displays.
The Priming option can be enabled at the time the Alaris
System is configured for use. The Priming selection (
soft key) is available only after the syringe type and medication
selection (prior to infusion mode selection).
3-8
Programming
Preparing Infusion
soft key.
CONFIRM

Priming

(Continued)
ALL SYRINGES
®
PRIME
Alaris
A
Syringe Selection
IMS Pre-fill 30 mL
>Select Syringe Size
ALL
CONFIRM
SYRINGES
WARNING
When priming:
• Ensure patient is not connected.
• Ensure air is expelled from
line prior to beginning infusion
(unexpelled air in line could have
serious consequences).
Failure to prime correctly can delay
infusion delivery and cause the total
volume to be infused to read higher
than the actual total delivered to the
patient.
CAUTION
During priming, the pressure limit
alarms are temporarily increased to
their maximum level.
®
System (with v9 Model 8000) DFU
PCA Module Section