Safety Clamp Fitment (
The primary administration set's safety clamp fitment is
a unique clamping device, on the pumping segment, that
prevents inadvertent free-flow when the administration set is
removed from the instrument.
Safety Clamp Fitment in Open Position
When a new Pump Module/Gemini Infusion System
administration set is removed from the package, the safety
clamp fitment is in the open position (white slide clamp aligned
with blue fitment). In this open position, flow is not occluded
but is allowed as required for the priming process. The roller
clamp is used to control flow during the priming process.
Safety Clamp Fitment in Closed Position
When a Pump Module/Gemini Infusion System administration
set is removed from the Pump Module, the instrument
automatically engages the safety clamp fitment in the closed
position (white slide clamp projects out from under blue
fitment). In this closed position, flow is occluded.
NOTE:
"Safety clamp" is referred to on the device as "Flo-Stop".
2-84
General Information
)
Pump Module
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
Blue Fitment
White Slide Clamp
(shown in open position)
Blue Fitment
White Slide Clamp
(shown in closed position)
Alaris
®
System (with v9 Model 8000) DFU