Symbols; Measurement Accuracy - Cardinal Health Alaris Directions For Use Manual

Supports guardrails suite mx with guardrails point-of-care software and v9 operating system software. with pc unit, model 8000.
Table of Contents

Advertisement

Specifications and Symbols (Continued)
See the PC Unit Section of this DFU for system symbols.
Silenced alarm.
Type BF equipment.
If the accuracy of any measurement does not seem
reasonable, first check the patient's vital signs by alternate
means and then check the
functioning properly.
An inaccurate measurement may be caused by:
Incorrect sensor application or use.
Significant levels of dysfunctional hemoglobins (such as
carboxyhemoglobin or methemoglobin).
Intravascular dyes (such as indocyanine green or
methylene blue).
Exposure to excessive illumination; such as, a surgical
lamp (especially one with a xenon light source), bilirubin
lamp, fluorescent light, infrared heating lamp or direct
sunlight.
Prolonged and/or excessive patient movement.
Venous pulsations.
Sensor placed on an extremity with a blood pressure
cuff, arterial catheter, intravascular line or other causes of
insufficient perfusion.
Nail aberrations, nail polish, fungus, etc. Remove nail
polish and/or move sensor to an unaffected site.
Placement is too close to electrosurgery equipment.
Defibrillation.
®
Alaris
System (with v9 Model 8000) DFU
SpO
Modules Section
2

Symbols

Measurement Accuracy

p
Module to ensure it is
S
O
2
4-27
General Information

Advertisement

Table of Contents
loading

Table of Contents