Cardinal Health Alaris Directions For Use Manual page 139

Supports guardrails suite mx with guardrails point-of-care software and v9 operating system software. with pc unit, model 8000.
Table of Contents

Advertisement

Syringe Module (Continued)
Pressure Sensing Disc Installed
(Continued)
Press Pressure Limit soft key.
4.
To enter a new pressure limit value, press Change Value
5.
soft key.
OR
If Auto Pressure feature is enabled, press Auto Pressure
soft key.
6.
Verify correct pressure limit input and press
key.
Pressure Sensing Disc NOT Installed
CHANNEL SELECT
1.
Press
OPTIONS
2.
Press
Alaris
®
System (with v9 Model 8000) DFU
Pump and Syringe Modules Section
Selecting Pressure Limit (Continued)
key.
key.
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
CONFIRM
soft
A
Channel Options 1 of 2
Guardrails Drugs
Guardrails IV Fluids
Multidose
Pressure Limit - 550 mmHg
Prime Set with Syringe
>Select an Option or EXIT
PAGE
EXIT
DOWN
A
Pressure Limit Selection
Current pressure
30
mmHg
Auto
Current limit
Pressure
_
550
Change
mmHg
value
CONFIRM
Programming
2-75