Messages - Cardinal Health Alaris Directions For Use Manual

Supports guardrails suite mx with guardrails point-of-care software and v9 operating system software. with pc unit, model 8000.
Table of Contents

Advertisement

Error
Hardware Detected Error
Missing Battery
Power Supply Error
System Error
Message
Battery Run Time = X.X hours
Low Battery
Panel Locked
Panel Unlocked
Powering Down
1-46
Troubleshooting and Maintenance
Alarms, Errors, Messages (Continued)
Errors (Continued)
Error detected on PC Unit.
Operation stops on all modules.
Battery not present or not
connected.
Power supply system malfunction.
Error detected on PC Unit.
Operation continues on all attached
modules.

Messages

power cord is disconnected
AC
from power source. Approximate
remaining battery run time under
current power consumption rate is
displayed.
Low battery threshold sensed;
remaining battery run time is
limited.
Tamper Resist feature is active and
key was pressed.
Tamper Resist feature deactivated.
Last module powering off. System
shuts off in indicated number of
seconds.
Meaning
Meaning
Alaris
Response
Replace PC Unit with an
operational instrument.
To continue temporary operation,
press
key. Replace
SILENCE
PC Unit with an operational
instrument.
Disconnect
power immediately.
AC
To continue operation under
battery power, press
SILENCE
key. Replace PC Unit with an
operational instrument.
To continue temporary operation,
press
key. Replace
SILENCE
PC Unit with an operational
instrument.
Response
Connect
power cord to power
AC
source as soon as possible.
Connect
power cord to power
AC
source (alarm silenced).
If appropriate, deactivate Tamper
Resist feature using Tamper Resist
Control on back of PC Unit.
None.
Press any key, except
key, to cancel power down
ON
sequence.
®
System (with v9 Model 8000) DFU
PC Unit Section
SYSTEM