Inspection Requirements - Cardinal Health Alaris Directions For Use Manual

Supports guardrails suite mx with guardrails point-of-care software and v9 operating system software. with pc unit, model 8000.
Table of Contents

Advertisement

Messages (Continued)
Model 8210 (Continued)
Message
Check Sensor - Motion Interference
Check Sensor - Weak Pulse
Check Sensor - Weak Signal
Model 8220
Message
Check Sensor - Low Perfusion
Check Sensor - Low Signal I.Q.
To ensure the Alaris
condition, both regular and preventive maintenance inspections
are required. Reference the System Maintenance software for
detailed instructions.
REGULAR INSPECTIONS
PROCEDURE
INSPECT FOR DAMAGE:
Case
IUI Connector
Keypad
CLEANING
START-UP
4-32
Troubleshooting and Maintenance
Alarms and Messages (Continued)
Patient's motion has inhibited
monitoring.
Patient's low perfusion has
inhibited monitoring.
Low quality of signal being
measured.
Patient's low perfusion has
inhibited monitoring.
Low signal quality being measured.

Inspection Requirements

®
System remains in good operating
FREQUENCY
Each usage
Each usage
Each usage
As required
Each usage
(8210)
Meaning
(8220)
Meaning
Failure to perform these
inspections may result in improper
instrument operation.
Regular and preventive
maintenance inspections should
only be performed by qualified
service personnel.
Alaris
Response
Check sensor. Move sensor to a
site with less motion.
Check sensor. Move sensor to a
better perfused site.
Check sensor. Confirm correct
sensor placement. Move sensor to
a better perfused site.
Response
Check sensor. Move sensor to a
better perfused site.
Check sensor. Confirm correct
sensor placement. Move sensor to
a better perfused site.
WARNING
CAUTION
®
System (with v9 Model 8000) DFU
SpO
Modules Section
2