Setting Bolus Dose (Continued)
CONFIRM
5.
Press
•
If programmed bolus dose is outside Soft Limit for that
care area, an audio alert sounds and a visual prompt
appears before programming can continue. If Yes soft
key is pressed, programming continues; if No soft key
is pressed, infusion needs to be reprogrammed.
•
If programmed bolus dose is outside Hard Limit for
that care area, an audio alert sounds and a visual
prompt appears before programming can continue.
Infusion needs to be reprogrammed.
If a Soft Limit is overridden, G icon displays. When G
•
soft key is pressed, all applicable out-of-range limits
are listed.
6.
If Authorization Code is disabled, door must be locked
prior to starting bolus dose.
7.
Verify dose value is correct and then press
key:
Delivering Bolus
•
•
Bolus and drug name alternate with VTBI in Main
Display
•
When bolus dose is complete,
scrolls in Channel Message Display.
•
Programmed infusion resumes.
NOTE:
BOLUS DOSE
The
has begun in
PCA
continuous infusion modes.
®
Alaris
System (with v9 Model 8000) DFU
PCA Module Section
Infusion Modes (Continued)
soft key.
scrolls in Channel Message Display
BOLUS COMPLETE
soft key is only available once an infusion
dose only, continuous infusion, or
START
soft
+
PCA
Guardrails Drug Setup
A
IMS Pre-fill 30mL
2
Cont. Dose:
mg/h
30
Max Limit:
mg/4 h
Bolus
1
mg
Dose:
[Conc]: 1 mg/mL
>Press START
PRO-
BOLUS
PAUSE
START
GRAM
DOSE
Programming
3-21