Cardinal Health Alaris Directions For Use Manual page 91

Supports guardrails suite mx with guardrails point-of-care software and v9 operating system software. with pc unit, model 8000.
Table of Contents

Advertisement

Primary Infusion
Continuous Infusion (Continued)
5.
An optional hospital-defined and editable starting value for
continuous infusion dose may already be entered.
OR
To make a rate or dose entry, press applicable soft key,
RATE
DOSE,
or
and use numeric data entry keys (other
value is calculated and displayed).
To enter volume to be infused, press VTBI soft key and
6.
use numeric data entry keys.
BOLUS
soft key appears only if Bolus Dose is enabled
within selected profile, drug is bolusable, and a
entered.
Syringe Module:
VTBI ALL
When
inactive and estimated available volume in syringe
displays. When
value can be entered.
VTBI ALL
When
pressed, estimated available volume in syringe
displays.
®
Alaris
System (with v9 Model 8000) DFU
Pump and Syringe Modules Section
- With Guardrails
Suite MX Protection (Continued)
®
VTBI
is enabled,
soft key is
VTBI
soft key is pressed, a new
VTBI
is disabled and
OR
is
VTBI
soft key is
FU
FU
RATE (mL/h)
RATE (mL/h)
CHANNEL
SELECT
CHANNEL
SELECT
PAUSE
PAUSE
CHANNEL
OFF
CHANNEL
OFF
RESTART
RESTART
Alteplase
A A A
Stroke
CONTINUOUS INFUSION
_ _ _
mL/h
RATE
_ _ _ _
mL
VTBI
_ _ _ _ _
DOSE
mg/kg/h
[Conc]: 1 mg/mL
>Select Rate or Dose
SETUP
BOLUS
Alteplase
A A A
Stroke
CONTINUOUS INFUSION
56.7
RATE
mL/h
_ _ _ _
VTBI
mL
0.81
_
mg/kg/h
DOSE
[Conc]: 1 mg/mL
>Select VTBI
SETUP
BOLUS
Alteplase
A A A
Stroke
CONTINUOUS INFUSION
56.7
RATE
mL/h
Available:
ALL
VTBI
(55.5 mL)
0.81
_
DOSE
mg/kg/h
[Conc]: 1 mg/mL
>Select VTBI
PAUSE
SETUP BOLUS
START
Programming
2-27

Advertisement

Table of Contents
loading