Cardinal Health Alaris Directions For Use Manual page 78

Supports guardrails suite mx with guardrails point-of-care software and v9 operating system software. with pc unit, model 8000.
Table of Contents

Advertisement

Preparing Syringe and Administration Set (
Priming - Using Options Menu (continued)
Administration Set With Pressure Sensing
Disc (Continued)
PRIME
7.
Press and hold
8.
Gently massage pressure sensing disc to ensure all air
is expelled. Disc must remain inverted only until air is
expelled. Continue to gently massage disc throughout
priming to ensure it does not become under- or over-filled.
9.
Continue to prime until fluid flows and priming is complete.
10. When priming is complete, release pressure sensing disc
PRIME
soft key.
and
2-14
Getting Started
soft key.
)
Syringe Module
(Continued)
®
Alaris
System (with v9 Model 8000) DFU
Pump and Syringe Modules Section
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
A
Prime Set with Syringe
CAUTION: Do
not prime with
disposable set
connected to
patient!
Prime Volume
0
mL
>Press and Hold PRIME
EXIT
PRIME
A
Prime Set with Syringe
CAUTION: Do
not prime with
disposable set
connected to
patient!
Prime Volume
2 mL
>Press and Hold PRIME
EXIT
PRIME