Getting Started; Introduction - Cardinal Health Alaris Directions For Use Manual

Supports guardrails suite mx with guardrails point-of-care software and v9 operating system software. with pc unit, model 8000.
Table of Contents

Advertisement

This Section of the DFU provides
instructions and information. It is used in conjunction with:
t
Module Technical Service Manual
E
CO
2
®
Microstream
Disposable Compatibility Card
Oridion's Microstream
PC Unit Section of this DFU
System Maintenance software (and its instructions) for Alaris
System check-in, maintenance, and wireless configuration
The
t
Module is a capnograph indicated for continuous,
E
CO
2
noninvasive monitoring of end tidal carbon dioxide (
fractional inspired carbon dioxide (
(
). The
t
Module and disposables are indicated for use
RR
E
CO
2
with intubated and nonintubated adult, pediatric and neonatal
patients. It is not intended for direct connection to ventilator
or breathing systems. Only one (
connected to the Alaris
The
t
Module is used with Oridion's patented
E
CO
2
®
Microstream
Disposables/circuits for sidestream capnography.
®
Microstream
Disposable: See "General Information" for
"Microstream
®
Disposable" Information.
Alarms and Messages: See "Troubleshooting and
Maintenance" for module-specific alarms and messages.
Contraindications: None known.
Electromagnetic Environment: See "Appendix" Section of
this DFU ("Regulations and Standards", "Compliance").
®
Alaris
System (with v9 Model 8000) DFU
EtCO
Module Section
2

Introduction

t
Module (Model
E
CO
2
®
disposable instructions
i
) and respiratory rate
F
CO
2
)
t
Module can be
1
E
CO
2
®
System.
)
8300
WARNING
Read all instructions, for both the
t
E
CO
2
using the Alaris
CAUTION
®
t
),
E
CO
2

Getting Started

Module and PC Unit, before
®
System.
O
nly
Getting Started
5-1