Supports guardrails suite mx with guardrails point-of-care software and v9 operating system software. with pc unit, model 8000.
Table of Contents

Advertisement

Preparing Syringe and Administration Set (
Loading (Continued)
2.
Open syringe barrel clamp.
a.
Pull syringe barrel clamp out and hold.
b.
Rotate clamp to left (clockwise or counter clockwise)
until it clears syringe chamber.
c.
Gently release clamp.
3.
Raise drive head to its fully extended position.
a.
Twist gripper control clockwise and hold in position.
b.
While holding gripper control in open position, raise
drive head to full extension.
c.
Gently release gripper control.
4.
Insert syringe (from front of instrument) by sliding flat edge
of syringe barrel flange between barrel flange grippers.
2-8
Getting Started
)
Syringe Module
(Continued)
Syringe Barrel Clamp Open
Drive Head Fully Extended
Gripper Control / Drive
Head Release in Open
Position
Plunger Grippers Open
®