Cardinal Health Alaris Directions For Use Manual page 243

Supports guardrails suite mx with guardrails point-of-care software and v9 operating system software. with pc unit, model 8000.
Table of Contents

Advertisement

Specifications and Symbols
Specifications (Continued)
Flow Rate Programming:
Rate Restriction by Syringe Size:
Fluid Ingress Protection:
Loading Dose Range:
Max Limit Range:
Occlusion Alarm Thresholds: Three settings:
Operating Principle:
PCA Dose Range:
Rate Accuracy:
Shock Protection:
®
Alaris
System (with v9 Model 8000) DFU
PCA Module Section
The flow rate range is from
Flow Rates (mL)
-
0.10
9.99
-
10
99.9
-
100
999
Syringe Size (mL)
/
50
60
/
30
35
20
IPX1, Drip Proof
Configured according to hospital best-practice guidelines.
Configured according to hospital best-practice guidelines.
Low
Medium
High
Positive displacement
Configured according to hospital best-practice guidelines.
of full-scale plunger travel (not including syringe variation)
±2%
WARNING
Syringe size and running force, variations of back pressure, or any combination
of these may affect rate accuracy. Factors that can influence back pressure
are: administration set configuration,
temperature. Back pressure may also be affected by type of catheter. See
"Trumpet and Start-Up Curves" for data on how these factors influence rate
accuracy.
Type
, Defibrillator Proof (PCA Module)
CF
Type
, Defibrillator Proof (Dose Request Cord)
BF
(Continued)
to
mL/h as follows:
0.1
999
Selectable Increments (mL/h)
0.01
0.1
1.0
Flow Rate Range (mL/h)
-
0.1
999
-
0.1
650
-
0.1
500
solution viscosity, and
IV
solution
IV
3-51
General Information