Errors; Messages - Cardinal Health Alaris Directions For Use Manual

Supports guardrails suite mx with guardrails point-of-care software and v9 operating system software. with pc unit, model 8000.
Table of Contents

Advertisement

Error
Channel Error
Syringe Calibration Required
Syringe Driver Head Error
Message
Bolus Complete
Infusion Complete
Load Complete
Max Limit Reached
3-58
Troubleshooting and Maintenance
Alarms, Errors, Messages (Continued)

Errors

Error detected. Operation stops on
affected module.
Error on infusing module indicating
calibration is required. Infusion
stops on affected module.
CALIBRATE
Display.
Noninfusing module, with plunger
grippers open, senses excessive
pressure being applied downward
on Drive Head.
in Message Display.

Messages

Current bolus dose completed.
Channel running in continuous
dose if programmed.
Current infusion completed.
Current loading dose completed.
Infusion mode menu available or
programmed infusion running.
Programmed maximum limit has
been reached over time period
specified. Infusion paused until
time limit has expired.
Meaning
scrolls in Message
OCCLUSION
scrolls
Meaning
Alaris
Response
To silence alarm and continue
operation of unaffected modules,
CONFIRM
press
soft key. Replace
module with an operational
instrument, as needed.
To silence alarm and continue
operation of unaffected modules,
CONFIRM
press
soft key. Replace
module with an operational
instrument, as needed.
To silence alarm and continue
normal operation, press
soft key.
Response
None
Set up a new infusion or press
CHANNEL OFF
key.
None
To silence alarm, press
key. To change Max Limit,
CHANNEL SELECT
press
PROGRAM
soft key, and unlock
door or enter Authorization Code
applicable for current Security
Access Level.
®
System (with v9 Model 8000) DFU
PCA Module Section
CONFIRM
SILENCE
, press