Disabling Security Access Code
The security code may be disabled for a specific infusion by
using the following procedure:
CHANNEL SELECT
1.
Press
2.
From Main Display, press
Press Security Code Access soft key.
3
DISABLE CODE soft k
4.
Press
CONFIRM
5.
Press
Security access code remains disabled until New
•
Patient?, Yes is selected in infusion startup or when
instrument remains powered off for more than
PAUSE
1.
Press
key.
OR
-- Continued on Next Page --
3-28
Programming
Infusion Modes (Continued)
key.
OPTIONS
key.
ey.
soft key.
Pausing Infusion
hours.
8
®
Alaris
System (with v9 Model 8000) DFU
A
Channel Options
2 of 2
Pressure Limit -
High
Prime Set with Syringe
Security Code Access
>Select an Option or
EXIT
PAGE
EXIT
UP
A
Security Code Access
Press DISABLE CODE to
temporarily turn off code access
capabilities for current patient.
Disable
Code
>Press DISABLE CODE or
EXIT
EXIT
CONFIRM
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
PCA Module
PCA Module Section