Cardinal Health Alaris Directions For Use Manual page 153

Supports guardrails suite mx with guardrails point-of-care software and v9 operating system software. with pc unit, model 8000.
Table of Contents

Advertisement

Features and Definitions (Continued)
Pump Module
Auto-Restart
Default Occlusion Pressure
Free Flow Protection
KVO Rate Adjust
Alaris
®
System (with v9 Model 8000) DFU
Pump and Syringe Modules Section
Features and Displays (Continued)
Part of Alaris
designed to minimize nuisance, patient-side occlusion alarms. Allows
system to automatically continue an infusion following detection of a
patient-side occlusion if downstream pressure falls to an acceptable
level within a
enabled, "Checking Line" function occurs when downstream pressure
exceeds pressure limit.
• In Selectable Pressure Mode: Pressure limit is either user-
adjustable or "locked" in system configuration.
• In Pump Pressure Mode: Pressure limit is a function of flow rate
and is automatically determined by device.
If downstream pressure decreases to a predetermined level, (below
of pressure limit) during
50%
infusion automatically continues. If condition is not cleared within
seconds, a "Partial Occlusion - Patient Side" alarm occurs.
15
Using Editor Software, system can be configured to allow
to
restart attempts within a rolling
9
number of restarts is exceeded or if feature is set to zero, an
"Occluded - Patient Side" alarm occurs when system detects
downstream pressure that exceeds pressure limit.
Starting occlusion pressure limit which can be configured by profile in
mmHg increments.
25
All Pump Module/Gemini Infusion System administration sets utilize
a unique clamping device (safety clamp on pumping mechanism)
to prevent inadvertent free-flow when administration set is removed
from instrument.
NOTE:
"Safety clamp" is referred to on the device as "Flo-Stop".
Used to select
which is rate of fluid flow after an "Infusion Complete" occurs.
rate never exceeds infusion rate.
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
®
System's Downstream Occlusion Detection system
-second "Checking Line" period. If this feature is
15
-second "Checking Line" period,
15
(Keep Vein Open) rate (
KVO
minute period. If allowable
10
to
mL/h allowed),
0.1
20
General Information
(zero)
0
KVO
2-89