Setting Up Continuous Infusion Only (Continued)
5.
To enter loading dose, press
Yes soft key and use numeric data entry keys.
6.
Verify parameters are correct and press
key.
•
If programmed parameters are outside Soft Limit for
that care area, an audio alert sounds and a visual
prompt appears before programming can continue.
If Yes soft key is pressed, programming continues;
if No soft key is pressed, infusion needs to be
reprogrammed.
•
If programmed parameters are outside Hard Limit for
that care area, an audio alert sounds and a visual
prompt appears before programming can continue.
Infusion needs to be reprogrammed.
If a Soft Limit is overridden, G icon displays. When G
•
soft key is pressed, all applicable out-of-range limits
are listed.
7.
Close and lock security door.
8.
Verify programming parameters are correct and press
START
soft key.
•
Green Infusing Status Indicator illuminates.
•
Infusion mode and drug name scroll in Channel
Message Display. If a loading dose has been entered,
DELIVERING LOAD
•
Volume infused in mL/h in Rate Display.
•
Main Display alternates between volume remaining
and infusion mode with drug name.
NOTES:
Time (in hours) associated with Max Limit is automatically
entered based on setup in system configuration.
Loading dose is included in volume infused but is not included
in Max Limit.
3-16
Programming
Infusion Modes (Continued)
LOAD DOSE
soft key, press
CONFIRM
scrolls
.
soft
®
Alaris
System (with v9 Model 8000) DFU
PCA Module Section
A
Continuous
Morphine
2
CONT
mg/h
DOSE
30
MAX
mg/4 h
LIMIT
_ _ 1
LOADING
mg
DOSE
[Conc]: 1 mg/mL
>Close and Lock Door Then
Press CONFIRM
CHANGE
CONFIRM
MODE
A
Second Nurse Summary
Morphine
2
Cont. Dose:
mg/h
30
Max Limit:
mg/4 h
Loading
1
mg
Dose:
[Conc]: 1 mg/mL
>Press START
PRO-
START
PAUSE
GRAM