Configurable Settings; Models 8210 And 8220; Model 8210 - Cardinal Health Alaris Directions For Use Manual

Supports guardrails suite mx with guardrails point-of-care software and v9 operating system software. with pc unit, model 8000.
Table of Contents

Advertisement

See the PC Unit Section of this DFU for system configurable
settings.
If the configuration settings need to be changed from the
Factory default settings, reference the applicable Technical
Service Manual or contact Cardinal Health Technical Support,
for technical, troubleshooting, and preventive maintenance
information.
With the Profiles feature enabled, the settings are configured
independently for each profile. A hospital-defined, best-
practice Data Set must be uploaded to enable the Profiles
feature. Date and Time is a system setting and is the same in
all profiles.

Models 8210 and 8220

Feature
Limit Mode
Pulse Beep Volume
Pulse Rate Alarm Limit, High
Pulse Rate Alarm Limit, Low
p
Alarm Limit, High
S
O
2
p
Alarm Limit, Low
S
O
2
Feature
SatSeconds
®
Alaris
System (with v9 Model 8000) DFU
SpO
Modules Section
2

Configurable Settings

Default Setting
Adult
1
Adult Mode:
Neonatal Mode:
Adult Mode:
Neonatal Mode:
Adult: Off
Neonatal:
Adult:
90%
Neonatal:

Model 8210

Default Setting
Off
Options
Adult, Neonatal
1
bpm
120
31
bpm
200
bpm
50
30
bpm
100
21
95%
20
80%
Options
,
10
seconds; Off
,
,
, Off
2
3
-
240
bpm
-
bpm
239
-
, Off
100%
-
99%
(8210)
,
,
25
50
100
General Information