Preparing Administration Set (
1.
Close roller clamp.
2.
Open Pump Module door.
•
Set's safety clamp fitment automatically closes to
prevent accidental free-flow.
3.
Remove set, as follows:
a.
Gently pull tubing below Air-in-Line Detector forward
and out.
b.
Lift upper fitment from upper fitment receptacle.
4.
If set is being removed to begin a gravity flow:
a.
Depress blue ridged release tab on upper side of
safety clamp device.
b.
Slide white slide clamp into blue fitment (open
position).
c.
Adjust flow rate using set's roller clamp.
NOTE:
"Safety clamp" is referred to on the device as "Flo-Stop".
1.
Prepare primary solution container in accordance with
manufacturer's directions for use.
2.
Open administration set package, remove set, and close
roller clamp. (Reference set's directions for use.)
3.
Insert administration set spike into prepared fluid
container, following accepted hospital/facility procedure,
and hang container
®
Alaris
System (with v9 Model 8000) DFU
Pump and Syringe Modules Section
Removing
Priming
inches above Pump Module.
20
) (Continued)
Pump Module
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
White Slide Clamp
(shown in closed position)
Getting Started
2-5