Connecting Microstream
3.
Release door.
4.
Connect Microstream
site and manner are dependent on patient intubation
status and type of Microstream
(reference disposable's directions for use).
NOTE:
The gas inlet is located on the lower left corner of the
instrument and is marked with a gas inlet symbol (
In the presence of high oxygen or anesthesia concentrations, it
may be necessary to connect a gas scavenging system to the
t
Module.
E
CO
2
1.
Open gas inlet/outlet door by turning door
counterclockwise until gas outlet is clearly visible. Hold in
open position.
2.
Secure gas scavenger system tubing to
firmly pushing tubing into fitting on gas outlet.
3.
Release door.
NOTE:
The gas outlet is located on the lower right corner of the
instrument and is marked with a gas outlet symbol (
®
Alaris
System (with v9 Model 8000) DFU
EtCO
Module Section
2
®
Disposable to patient. Connection
®
Disposable being used
Attaching Gas Scavenging System
®
Disposable (Continued)
).
t
Module by
E
CO
2
).
Gas scavenging system
Getting Started
5-3