Messages - Cardinal Health Alaris Directions For Use Manual

Supports guardrails suite mx with guardrails point-of-care software and v9 operating system software. with pc unit, model 8000.
Table of Contents

Advertisement

Model 8210 (Continued)
Alarm
Check Sensor - Weak Pulse
Check Sensor - Weak Signal
Model 8220
Alarm
Check Sensor - Light
Check Sensor - Low Perfusion
Check Sensor - Low Signal I.Q.
Model 8210
Message
Check Sensor - Electrical or Optical
Interference
Check Sensor - High Pulse
Amplitude
Check Sensor - Excessive Ambient
Light
®
Alaris
System (with v9 Model 8000) DFU
SpO
Modules Section
2
Alarms and Messages (Continued)
Alarms (Continued)
Patient's low perfusion has
inhibited monitoring.
Low quality of signal being
measured.
Light interference on sensor.
Patient's low perfusion has
inhibited monitoring.
Low signal quality being measured.

Messages

External interference on sensor.
Artifact interfering with pulse
reading.
Light interference on sensor.
(8210)
Meaning
(8220)
Meaning
(8210)
Meaning
Troubleshooting and Maintenance
Response
Check sensor. Move sensor to a
better perfused site.
Check sensor. Confirm correct
sensor placement. Move sensor to
a better perfused site.
Response
Check sensor. Remove or reduce
lighting. Cover or reposition
sensor.
Check sensor. Move sensor to a
better perfused site.
Check sensor. Confirm correct
sensor placement. Move sensor to
a better perfused site.
Response
Check sensor. Identify source of
external interference if other than
sensor.
Check sensor. Relocate sensor to a
site with less artifact interference.
Check sensor. Remove or reduce
lighting. Cover or reposition
sensor.
4-31