Cardinal Health Alaris Directions For Use Manual page 251

Supports guardrails suite mx with guardrails point-of-care software and v9 operating system software. with pc unit, model 8000.
Table of Contents

Advertisement

Messages (Continued)
Message
NEOI (Near End of Infusion)
Panel Locked
Panel Unlocked
Pause
PCA Complete
PCA Not In Secure Location
Syringe Not Recognized
®
Alaris
System (with v9 Model 8000) DFU
PCA Module Section
Alarms, Errors, Messages (Continued)
Syringe almost empty.
Tamper Resist feature is active and
a key was pressed.
Tamper Resist feature deactivated.
Pause control pressed; infusion
stopped.
Current
Channel running in continuous
dose if programmed.
PCA Module is not in preferable
location to allow locking to PC Unit.
Device is not in a tamper evident
position.
Installed syringe of unknown type
and size.
Meaning
dose complete.
PCA
Response
This is a timed event that can be
set or changed (see "General
Information", "Configurable
Settings").
To silence alarm, press
key. PCA Module remains
functional and continues infusion.
Green indicator light is lit and
yellow light flashes. PCA Module
is silent until Syringe Empty alarm
sounds (see Syringe Empty alarm
response).
If appropriate, deactivate Tamper
Resist feature using Tamper Resist
Control on back of PC Unit.
None.
To resume infusion, press
CHANNEL SELECT
key, or press
START
key and then
None.
Detach PCA Module from current
position and reattach to immediate
right of PC Unit.
Select and confirm correct
syringe type and size, and then
CONFIRM
press
, or use a syringe
type and size that system can
automatically and correctly identify.
Troubleshooting and Maintenance
SILENCE
RESTART
soft key.
3-59