4.
When priming is complete, release
5.
To return to main screen, press
Guardrails Drug Setup screen displays.
6.
Select infusion mode
NOTE:
Fluid is delivered during priming only while the PRIME soft
key is pressed. Each press of the PRIME soft key delivers
up to 2 mL of priming/fluid per continuous press. To deliver
additional amounts, press the PRIME soft key again.

Programming an Infusion

1.
Perform steps in "Getting Started", "Preparing Syringe and
Administration Set".
2.
Perform following steps (see PC Unit Section of this DFU,
"General Setup and Operation", "Startup"):
a.
Power on system.
Choose Yes or No to New Patient?
b.
c.
Select profile, if required.
d.
Enter patient identifier, if required.
CHANNEL SELECT
3.
Press
4.
Unlock security door or set key to
5.
Confirm time of day or change time if necessary.
3-10
Programming
Preparing Infusion (Continued)
Priming (Continued)
PRIME
EXIT
.
key.
PROGRAM
soft key.
soft key.
position.
A
PCA
Time of Day
Current time:
08:00
>CONFIRM Time-of-Day
EXIT
®
Alaris
System (with v9 Model 8000) DFU
PCA Module Section
Change
Time
CONFIRM