Bolus Dose (Continued)
8.
Syringe Module: If bolus dose was programmed at
beginning of infusion, attach administration set to
patient.
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
Stopping Bolus Dose
The display examples in this procedure represent stopping a
bolus dose which was programmed using the Drug Library.
Even where the displays are different when stopping a bolus
dose which was programmed using a non-library drug, the
procedure is the same.
CHANNEL SELECT
1.
Press
STOP BOLUS
2.
Press
To stop bolus and start continuous infusion, press Yes soft
3.
key.
®
Alaris
System (with v9 Model 8000) DFU
Pump and Syringe Modules Section
- With Guardrails
®
key.
soft key.
Suite MX Protection (Continued)
FU
FU
RATE (mL/h)
RATE (mL/h)
CHANNEL
SELECT
CHANNEL
SELECT
PAUSE
PAUSE
CHANNEL
OFF
CHANNEL
OFF
RESTART
RESTART
Alteplase
A A A
Stroke
BOLUS DOSE
Dose
0.090
mg/kg
Remaining:
70
Pt. Weight:
kg
1
Time Left:
min
TOTAL DOSE = 6.3 mg
BOLUS VTBI = 6.3 mL
>Press START to Continue
Infusing Bolus Dose
STOP
PAUSE
START
BOLUS
Alteplase
A
Stroke
Stop Bolus and Start
Yes
Continuous infusion?
No
>Press Yes or No
Programming
2-33