Displays; Configurable Settings - Cardinal Health Alaris Directions For Use Manual

Supports guardrails suite mx with guardrails point-of-care software and v9 operating system software. with pc unit, model 8000.
Table of Contents

Advertisement

The displays illustrated throughout this document are for
illustration purposes only. The display content varies,
depending on configuration settings, type of administration
set in use, hospital-defined Data Set uploaded using
®
the Guardrails
Suite
parameters, and many other variables. Reference specific
drug product labeling for information concerning appropriate
administration techniques and dosages.
See the PC Unit Section of this DFU for system configurable
settings.
The configuration settings are selected during Data Set
development and then uploaded to the Alaris
of the data.
With the Profiles feature enabled, the settings are configured
independently for each profile. A hospital-defined, best-
practice Data Set must be uploaded to enable the Profiles
feature. Date and Time is a system setting and is the same in
all profiles.
Feature
Authorization Code
Bolus Delivery Rate
Bolus Dose
Bolus Dose include in Max. Limit
Dose Request Cord Configuration
Loading Dose
Lockout Interval
Max Accumulated Dose Range
Max Rate (for Continuous Dose)
3-48
General Information
Features and Displays (Continued)

Displays

, programmed drug calculation
MX

Configurable Settings

Default Setting
1 - 99 minutes in 1-minute
®
System as part
None
150 mL/h
Enabled
Disabled
Profile 2
Enabled
increments
1-hour limit
999 mL/h
Options
4 digits (0 - 9)
One code applies to all profiles
75 - 500 mL/h (limited by syringe size)
Enabled - Disabled
Enabled - Disabled
Profile 1, 2, 3
Enabled - Disabled
Min/Max 1 - 99 minutes
Disabled; 1, 2 or 4-hour limit
0.1 - 99.9 mL/h in 0.1 mL/h increments;
100 - 999 mL/h in 1 mL/h increments
®
Alaris
System (with v9 Model 8000) DFU
PCA Module Section