Errors - Cardinal Health Alaris Directions For Use Manual

Supports guardrails suite mx with guardrails point-of-care software and v9 operating system software. with pc unit, model 8000.
Table of Contents

Advertisement

Pump and Syringe Modules
Error
Channel Error
Syringe Module
Error
Syringe Calibration Required
Syringe Driver Head Error
Alaris
®
System (with v9 Model 8000) DFU
Pump and Syringe Modules Section
Alarms, Errors, Messages (Continued)

Errors

Error detected. Operation stops on
affected module.
Error on infusing module indicating
calibration is required. Infusion
stops on affected module.
CALIBRATE
Display.
Noninfusing module, with plunger
grippers open, senses excessive
pressure being applied downward
on Drive Head.
in Message Display.
FU
FU
RATE (mL/h)
RATE (mL/h)
CHANNEL
SELECT
CHANNEL
SELECT
PAUSE
PAUSE
CHANNEL
OFF
CHANNEL
OFF
RESTART
RESTART
Meaning
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
Meaning
scrolls in Message
OCCLUSION
scrolls
Troubleshooting and Maintenance
Response
To silence alarm and continue
operation of unaffected module(s),
CONFIRM
press
soft key. Replace
module with an operational
instrument, as needed.
Response
To silence alarm and continue
operation of unaffected module(s),
CONFIRM
press
soft key. Replace
module with an operational
instrument, as needed.
To silence alarm and continue
normal operation, press
soft key.
CONFIRM
2-121

Advertisement

Table of Contents
loading

Table of Contents