Cardinal Health Alaris Directions For Use Manual page 79

Supports guardrails suite mx with guardrails point-of-care software and v9 operating system software. with pc unit, model 8000.
Table of Contents

Advertisement

Preparing Syringe and Administration Set (
Priming - Using Options Menu (continued)
Administration Set With Pressure Sensing
Disc (Continued)
11. Reinstall pressure sensing disc, as follows:
a.
Orient pressure sensing disc, as follows:
fluid side up (patient side down)
cavity forward (membrane toward instrument)
b.
Gently slide pressure sensing disc up into slot in
pressure sensing disc housing.
c.
Apply firm upward pressure on pressure sensing disc
(not tubing) until disc snaps into place.
12. To return to main screen, press
EXIT
If
soft key is pressed before pressure sensing
disc is reinstalled, a prompt to reinstall pressure
sensing disc displays.
®
Alaris
System (with v9 Model 8000) DFU
Pump and Syringe Modules Section
Syringe Module
EXIT
soft key.
FU
RATE (mL/h)
)
(Continued)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
Syringe Module
Slot in Pressure
Fluid Side
Sensing Disc
Housing
Pressure
Sensing Disc
Patient Side
Pressure Sensing Disc
A A A
Re-install Pressure Sensing
Disc.
>Press CONFIRM
CONFIRM