Scheduling a Callback
When programming a Delay for or Delay until infusion, a
callback can be scheduled for that infusion. There are three
types of callback:
Before - gives an alert when delay period is completed and
•
infusion needs to be initiated.
After - gives an alert when delayed infusion has
•
completed.
Before and After - gives an alert when delay period is
•
completed and infusion needs to be initiated and when
delayed infusion has completed.
The default callback (None), or the callback for the current
profile, appears on the Main Display. To schedule a different
callback:
1.
Prior to pressing
Delay for or Delay until programming process, press
CALL BACK
soft key.
2.
Press soft key corresponding to desired callback option.
•
Scheduled callback appears on Main Display.
3.
To initiate delay, press
If Delay until programming, time infusion is scheduled
•
to start appears on Main Display.
OR
If Delay for programming, delay period counts down
on Main Display.
If Before option was selected:
•
♦
An audio prompt sounds when delay period has
ended.
♦
Yellow Standby Status Indicator flashes.
DELAY COMPLETE
♦
appears on Main Display.
-- Continued on Next Page --
2-66
Programming
Delay Options (Continued)
CONFIRM
soft key to initiate delay during
CONFIRM
soft key.
scrolls in Message Display and
A
Delay Options
Pause
Delay for
09:00
Delay until
CALL
> > > > > >
BACK
>Select a Callback Option
EXIT
®
Alaris
System (with v9 Model 8000) DFU
Pump and Syringe Modules Section
FU
FU
RATE (mL/h)
RATE (mL/h)
CHANNEL
SELECT
CHANNEL
SELECT
PAUSE
PAUSE
CHANNEL
OFF
CHANNEL
OFF
RESTART
RESTART
08:00
Before
Before
And After
After
None
CONFIRM