Programming an Infusion (Continued)
NOTES:
If the programmed "_ _ / _ _ mL" concentration is outside
the Soft Limit, an audio alert sounds and a visual prompt
appears before programming can continue. If the Yes soft
key is pressed, programming continues; if the No soft key is
pressed, the infusion must be reprogrammed.
If the programmed "_ _ / _ _ mL" concentration is outside the
Hard Limit for that care area, an audio alert sounds and a
visual prompt appears before programming can continue. The
drug amount and diluent volume must be reprogrammed.
If a Soft Limit is overridden, the G icon displays. When the G
soft key is pressed, all applicable out-of-range limits are listed.
Programming Parameters
The PCA Module uses the following programming parameters,
depending on infusion mode selected. See "General
Information", "Features and Definitions" for infusion mode
definitions and features.
PCA Dose: patient self-administered dose.
•
Lockout Interval: programmed time elapse between
•
availability of
PCA
Continuous Dose: basal rate dose.
•
Max Limit: (optional) total amount of drug which can be
•
infused over a specified time period.
Loading Dose: (optional) bolus dose infused prior to
•
initiation of
infusion.
PCA
Bolus Dose: (optional) additional dose programmed after
•
the initiation of
PCA
NOTE:
When the PC Unit is in the Infusion Mode Selection, Infusion
Setup or Bolus Setup screens, a patient dose request from the
Dose Request Cord is handled as an unmet demand.
3-12
Programming
Preparing Infusion (Continued)
Infusion Modes
doses.
infusion.
®
Alaris
System (with v9 Model 8000) DFU
PCA Module Section