Figure 72 Calibration Offsets
Graphical Report
Quantum Scalar i6000 User's Guide
are installed properly and that the calibration targets are correctly
snapped on to the magazines.
• If the graphical report shows a failed result for the I/E station, make
sure the I/E station and front door are completely shut.
• If running the test again still generates failed results, realign the
middle X-axis rail and check the alignment of the top and bottom X-
axis rails at the location of the failure.
See
Figure 72
on page 637.
Chapter 15: Maintaining Your Library
Maintenance Actions
637