3.2.5 Measure Mode
Set the measure mode to conductivity, salinity or TDS.
Press the or key to scroll through the Setup
Menu list until the Measure Mode item is shown.
Press the or key to set the measure mode.
Press the "Enter" key to save the change.
3.2.6 Nominal Cell Constant
Set the nominal value for the cell constant (K) of the
conductivity electrode being used.
The entered nominal cell constant (K) is used to determine
conductivity, salinity and TDS measurement values from
the conductance reading. Once a calibration is performed,
the entered nominal cell constant (K) multiplied by the
calculated calibration factor is used to determine
conductivity, salinity and TDS measurement values from
the conductance reading.
The entered nominal cell constant (K) is also used for
automatic conductivity standard recognition during
conductivity calibrations.
Press the or key to scroll through the Setup
Menu list until the Nominal Cell Constant item is
shown.
Press the or key to set the cell constant (K)
value.
a. Press the key to increase the value by
0.001. Press and hold the key to quickly
increase.
b. Press the key to decrease the value by
0.001. Press and hold the key to quickly
decrease.
Press the "Enter" key to save the change.
Eutech EC 1720 Conductivity Meter
17